Test Code LAB4893 Blood Gases, Cord Blood, Arterial
Additional Codes
Test Name in EPIC | EPIC Test Code | Mnemonic |
---|---|---|
CORD BLOOD GASES, ARTERIAL | LAB4893 | ART CORD GAS |
Methodology
pH: pH-sensitive glass electrode in conjunction with a reference electrode
pCO2: An electrode that measures pCO2 ions directly as a pH change due to the diffusion of CO2 across the membrane
PO2: Measured through a Clark-type electrode. Oxygen diffuses through the membrane and continuously replaces oxygen molecules in the electrolyte layer during a cathode reaction.
O2 Saturation: Calculated
HCO3: Calculated
Includes pH, pCO2, PO2, O2 saturation, HCO3, and base excess.
Specimen Requirements
Specimen must arrive within 30 minutes of draw.
Specimen Type: Cord blood, arterial
Container/Tube: Heparin blood gas syringe
Specimen Volume: 0.3 mL
Specimen Minimum Volume: 0.15 mL
Collection Instructions: Remove needle, express air bubbles, and cap with rubber stopper. Microtainers are not acceptable for specimen collection.
Additional Information: Label specimen appropriately (arterial cord blood).
Specimen Transport Temperature
Ambient 30 minutes/Refrigerate NO/Frozen NO
Day(s) Test Set Up
Monday through Sunday
Performing Laboratory
Chemistry -
East Region: Deer River Hospital Lab, SMDC Clinical Lab, Virginia Hospital Lab
West Region: Fargo Hospital, Fosston Hospital Lab, St. Mary's - Detroit Lakes Hospital Lab
Test Classification and CPT Coding
82803
Reference Values
No established reference values