Test Code LABBULLP Bullous Pemphigoid, BP180 and BP230, IgG Antibodies, Serum
Additional Codes
Test Name in EPIC | EPIC Test Code | Mnemonic | Mayo Test ID |
---|---|---|---|
Bullous Pemphigoid, 180 and 230 IgG Antibodies | LABBULLP | BPAB | BPAB |
Specimen Required
Collection Container/Tube:
Preferred: Red top
Acceptable: Serum gel
Submission container/tube: Plastic vial
Specimen Volume: 1 mL
Collection Instructions: Centrifuge and aliquot serum into a plastic vial.
Useful For
Initial screening test in the diagnosis of bullous pemphigoid and its variants
Complementing the standard serum test of indirect immunofluorescence utilizing primate esophagus substrate and primate salt-split skin substrate (CIFS / Cutaneous Immunofluorescence Antibodies [IgG], Serum)
Method Name
Enzyme-Linked Immunosorbent Assay (ELISA)
Reporting Name
BP 180 and 230, SerumSpecimen Type
Serum RedSpecimen Minimum Volume
0.5 mL
Specimen Stability Information
Specimen Type | Temperature | Time | Special Container |
---|---|---|---|
Serum Red | Refrigerated (preferred) | 14 days | |
Frozen | 30 days | ||
Ambient | 14 days |
Reject Due To
Gross hemolysis | OK |
Gross lipemia | OK |
Gross icterus | OK |
Clinical Information
Bullous pemphigoid (BP) is a chronic pruritic blistering disorder found mainly in aged persons, characterized by the development of tense blisters over an erythematous or urticarial base. IgG anti-basement membrane zone antibodies are found in the serum of patients, and linear IgG and complement component 3 sediment is found on the basement membrane zone of the lesion. Several well characterized variants exist including localized, mucous membrane predominant and pemphigoid gestationis, also referred to as herpes gestationis.
Target antigens of the autoantibodies in BP patient serum are BP230 and BP180, also called BPAG1 and BPAG2. Molecular weight of these antigens is 230 kDa and 180 kDa, respectively. BP180 is thought to be the direct target of the autoantibody because of its location along the basement membranes, and the autoantibody against BP230 is thought to be secondarily produced.
Reference Values
BULLOUS PEMPHIGOID 180:
<20 RU/mL (negative)
≥20 RU/mL (positive)
BULLOUS PEMPHIGOID 230:
<20 RU/mL (negative)
≥20 RU/mL (positive)
Interpretation
Antibodies to bullous pemphigoid (BP) BP180 and BP230 have been shown to be present in most patients with pemphigoid. Adequate sensitivities and specificity for disease are documented and Mayo Clinic's experience demonstrates a very good correlation between BP180 and BP230 results and the presence of pemphigoid (see Supportive Data). However, in those patients strongly suspected to have pemphigoid, either by clinical findings or by routine biopsyor direct immunofluorescence, and in whom the BP180/BP230 assay is negative, follow-up testing by CIFS / Cutaneous Immunofluorescence Antibodies (IgG), Serum is recommended.
Antibody titer may correlate with disease activity in some patients. Patients with severe disease may be expected to have high titers of antibodies to BP. Titers may decrease with clinical improvement.
Cautions
As with other diagnostic test procedures, the results obtained with bullous pemphigoid (BP) BP180 and BP230 enzyme-linked immunosorbent assay kit serve only as an aid to diagnosis and should not be interpreted as diagnostic in themselves.
Performance of the assay in pediatric patients has not been established.
Performance of the assay on other matrices besides serum has not been established.
Supportive Data
Thirty-two classic bullous pemphigoid (BP), 15 mucous membrane pemphigoid, and 7 other pemphigoid variants, diagnosed by direct immunofluorescence, routine histology, and clinical presentation were tested. Controls included 47 patients with other autoimmune blistering disorders and 42 age-matched controls without skin disease. Forty of 54 (74%) patients with BP and variants tested positive for BP180 and/or BP230 autoantibodies. Of these patients, 28 of 32 (88%) with classical BP, 8 of 15 (53%) with mucous membrane predominant (MMP), and 4 of 7 (57%) of other pemphigoid variants, tested positive.
The calculated sensitivities in classical BP were 54% for BP180 alone and 56% for BP230 alone. The sensitivity increased to 88% with both tests combined, which is comparable to that of indirect immunofluorescence (IIF) (88%). In MMP the calculated sensitivities were 47% for BP180 alone, 13% for BP230 alone, and 53% for both combined. This was slightly less than the sensitivity of IIF (67%). Only 5 of 47 (11%) and 2 of 47 (4%) control patients with other autoimmune blistering disorders were positive for BP180 and BP230 autoantibodies respectively. Interestingly, the 2 patients positive for BP230 autoantibody had paraneoplastic pemphigus. One of 42 (2%) and 0 normal controls tested positive for BP180 and BP230 respectively.
The calculated specificities for BP180, BP230, and IIF were 93%, 98%, and 92% respectively.
Method Description
This enzyme-linked immunosorbent assay (ELISA) method detects and measures serum levels of antibodies of certain pemphigoid diseases. Calibrators and patient sera are added to microwells coated with bullous pemphigoid (BP) BP180 and BP230 antigens, allowing antibodies to react with the immobilized antigens. After washing to remove any unbound serum proteins, horseradish peroxidase-conjugated IgG is added and incubated. Following another wash step, the peroxidase substrate is added and allowed to incubate for an additional period. Stop solution is then added to each well to cancel the enzyme reaction and to stabilize the color development. The assay can be quantified by measuring the reaction photometrically and plotting the results. The amount of antigen specific bound antibody is proportional to the color intensity.(Package inserts: Anti-BP180-NC16A-4X ELISA [IgG], Form EA_1502-2G_A_US_D04. EuroImmun; 07/08/2020; Anti-BP230-CF ELISA [IgG], Form EA_1502-1G_A_UK_C03.doc. EuroImmun; 10/23/2020)
Day(s) Performed
Varies
Report Available
1 to 5 daysPerforming Laboratory
Mayo Clinic Laboratories in RochesterTest Classification
This test has been modified from the manufacturer's instructions. Its performance characteristics were determined by Mayo Clinic in a manner consistent with CLIA requirements. This test has not been cleared or approved by the US Food and Drug Administration.CPT Code Information
83516 x 2
LOINC Code Information
Test ID | Test Order Name | Order LOINC Value |
---|---|---|
BPAB | BP 180 and 230, Serum | 92671-7 |
Result ID | Test Result Name | Result LOINC Value |
---|---|---|
606816 | BP 180, S | 53842-1 |
606817 | BP 230, S | 53843-9 |