Sign in →

Test Code LABCBGV Blood Gases, Cord Blood, Venous

Additional Codes

Test Name in EPIC EPIC Test Code Mnemonic
CORD BLOOD GASES, VENOUS LABCBGV VEN CORD GAS

 

Methodology

pH: pH-sensitive glass electrode in conjunction with a reference electrode

pCO2: An electrode that measures pCO2 ions directly as a pH change due to the diffusion of CO2 across the membrane

PO2: Measured through a Clark-type electrode. Oxygen diffuses through the membrane and continuously replaces oxygen molecules in the electrolyte layer during a cathode reaction.

O2 Saturation: Calculated

HCO3:  Calculated

Includes pH, pCO2, PO2, O2 saturation, HCO3, and base excess.

Specimen Requirements

Specimen must arrive within 30 minutes of draw.

 

Specimen Type: Cord blood, venous

Container/Tube: Heparin blood gas syringe

Specimen Volume: 0.3 mL

Specimen Minimum Volume: 0.15 mL

Collection Instructions: Remove needle, express air bubbles, and cap with rubber stopper. Microtainers are not acceptable for specimen collection.

Additional Information: Label specimen appropriately (venous cord blood).

Specimen Transport Temperature

Ambient 30 minutes/Refrigerate NO/Frozen NO

Day(s) Test Set Up

Monday through Sunday

Performing Laboratory

Chemistry -

East Region: Deer River Hospital Lab, SMDC Clinical Lab, Virginia Hospital Lab

West Region: Fargo Hospital, Fosston Hospital Lab, St. Mary's - Detroit Lakes Hospital Lab

 

Test Classification and CPT Coding

82803

Reference Values

No established reference values