Test Code LABTHRTM Thrombin Time, Plasma
Additional Codes
Test Name in EPIC | EPIC Test Code | Mnemonic |
---|---|---|
THROMBIN TIME, PLASMA | LABTHRTM | THRTM |
Methodology
Electromechanical Clot Detection
Specimen Requirements
Specimen Type: Platelet-poor plasma
Container/Tube:
Fill Blue top tube to the etched line located just below the cap of the tube.
Preferred: Full Light-blue top (3.2% sodium citrate)
Acceptable: Full Sodium citrate (BD® or Greiner tubes)
Specimen Volume: 1 mL
Specimen Minimum Volume: 0.5 mL
Collection Instructions:
1. Unacceptable specimens: Specimens that are short-draws, over-draws, clotted or hemolyzed may yield incorrect results.
2. If sending from a site spin down, remove top 2/3 of plasma, and spin plasma again in a polyproyplene (cloudy plastic) vial.
3. Aliquot top 3/4 and place platelet-poor plasma in a polypropylene (cloudy plastic) vial.
4. Glass vials are not acceptable for processing/transport/storage.
Additional Information:
1. Double-centrifuged specimens are critical for accurate results as platelet contamination may cause spurious results.
2. Send specimen ambient. If there is a delay in transport of > 8 hours, send specimen frozen.
Specimen Transport Temperature
Refrigerate NO/Frozen OK/Ambient* 8 hours OK
*Ambient 8 hours OK when patient is not on heparin therapy
Ambient 2 hours OK when patient is on heparin therapy (either Unfractionated (UFH) or Low Molecular Weight (LMW)).
Day(s) Test Set Up
Monday through Sunday
Performing Laboratory
Coagulation-
SMDC Clinical Lab
Test Classification and CPT Coding
85670
Reference Values
Normal range changes every year with lot change of Thrombin Time reagent. See patient chart copy for current range.