Test Code LABTHRTM Thrombin Time, Plasma
Additional Codes
| Test Name in EPIC | EPIC Test Code | Mnemonic |
|---|---|---|
| THROMBIN TIME, PLASMA | LABTHRTM | THRTM |
Methodology
Electromechanical Clot Detection
Specimen Requirements
Fill Blue top tube to the etched line located just below the cap of the tube.
Grossly hemolyzed specimen is not acceptable.
Light-blue top (3.2% sodium citrate)
Specimen Transport Temperature
1. To achieve platelet-poor plasma, spin down, remove top 2/3 of plasma, and spin plasma again in a plastic vial.*
2. Aliquot top 3/4 and place platelet-poor plasma in a plastic transport vial.*
3. Glass vials are not acceptable for processing/transport/storage.
4. Preferred specimen volume: 1 mL sodium citrate plasma
5. Minimum specimen volume: 0.5 mL sodium citrate plasma
* 2 Aliquot labels will auto print at non-performing sites to double spin
8 hours at 20 ºC; If on heparin therapy, plasma stable for 2 hours at 20±5°C
1 month at -20 ºC. (Double spun)
Day(s) Test Set Up
Monday through Sunday
Performing Laboratory
Coagulation-
SMDC Clinical Lab
Test Classification and CPT Coding
85670
Reference Values
Normal range changes every year with lot change of Thrombin Time reagent. See patient chart copy for current range.